Power Components, Power Adapters, External Power Supplies, DC-DC Converters, Open Frame Power Supplies and more for industry
 
HPU10D SERIES
HPU10D SERIES

10W External power supplies for medical applications including home healthcare equipment,
patient monitors, blood pressure equipment, portable medical devices, ECG machines, many others


2 Year warranty.
FEATURES
Wide Operating Voltage, 80 to 275 VAC, 47 to 440 Hz AU Plug
Medical Safety 3rd (IEC60601-1 3rd Edition) Input to Output : 2MOPP
Single Output Energy Star 2.0, Efficiency Level V
Over Voltage and Over Load Protection Class II System
Electrical Characteristics:
Vins Safety Approvals Input Voltage Range Safety Approval & Specification in Label 100-240VAC
Vin Operate Voltage Range See derating curve 80-275VAC
Fi Input Frequency Sine wave 47-440Hz
Po Output Power Range See rating chart 10W
lil Input Current (Low Line) lo=Full load, Vin=100VAC 0.3A
lih Input Current (High Line) lo=Full load, Vin=240VAC 0.2A
Ir Low Line inrush Current lo=Full load, 25°C, Cool start, Vin=100VAC 40A (max.)
High Line Inrush Current lo=Full load, 25°C, Cool start, Vin=240VAC 100A (max.)
Eff Efficiency lo=Full Load, Vin=230VAC 67-85%
REG-i Line Regulation lo=Full Load, Vin=100-240VAC 1% (mAX.)
REG-o Load Regulation Vin=230VAC 3-5%
OVP Over Voltage Protection Nil
OCP Over Current Protection Nil.But,Output protected to short circuit conditions
Ttr Time of Transient Response lo=Full Load, Vin=110VAC 4mS (max.)
Thu Hold-up Time lo=Full Load, Vin=110VAC 12mS (typ.)
Ts Start Up Time lo=Full Load, Vin=100-240VAC 2S (max.)
VP-p Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC 1% (max.)
Tc Temperature Coefficient All Condition ±0.04%/°C
Pno No-Load Power Consumption No load, Vin=230VAC See rating chart
Vps Dialechic Withstanding Voltage Primary to secondary 6612VDC (min.)
EMI EMC Emission Compliance to EN55011 (CISPR11), EN61000-3-2,-3 B Class


Environmental:

ToOperating TemperatureSee derating curve-10-70°C
TsStorage Temperature10 — 95% RH-40-85°C
HoOperating HumidityNon-condensing0-95% RH
HrStorage Humidity See Rating Chart0-95% RH
VsgSurge VoltageAll Condition2kV (min.)
ESDaElectro Static DischargeAir Discharge, IEC61000-4-28kV (min.)
ESDcElectro Static DischargeContact Discharge, IEC61000-4-26kV (min.)
MTBFMean Time Between FailureOperating temperature at 25°C, Calculated per MIL-HDBK-217F0.1M Hrs (min.)
ELEVOperating Altitude (Elevation)All condition5000m (max.)
VBRVibration10 ~ 500Hz, 10min./1cycle,60min. Each along x, Y, Z axes.5G (min.)

Mechanical Specifications:




1. Operating Temperature: -10 to 70°C
2. Derate linearly from 100% load at 40°C to 50% load at 70°C
Output Voltage and Current Rating Chart ( Single Output )
Model No. Output
Voltage
Output Current Total Regulation Max. Output Power Pno (max)
HPU10D-101 3.3 VDC 1.81 A 5% 8W 0.2W
HPU10D-102 5.0 VDC 1.60 A 5% 8W 0.3W
HPU10D-103 6 ~ 7 VDC 1.33 ~ 1.14 A 5% 8W 0.3W
HPU10D-104 8 ~ 11 VDC 1.25 ~ 0.90 A 5% 6W 0.3W
HPU10D-105 12 ~ 13 VDC 0.83 ~ 0.76 A 5% 10W 0.3W
HPU10D-106 14 ~ 16 VDC 0.71 ~ 0.62 A 5% 10W 0.3W
HPU10D-107 17 ~ 20 VDC 0.58 ~ 0.50 A 5% 10W 0.3W
HPU10D-108 21 ~ 26 VDC 0.47 ~ 0.38 A 5% 10W 0.3W
HPU10D-109 27 ~ 32 VDC 0.37 ~ 0.31 A 3% 10W 0.3W
HPU10D-110 33 ~ 39 VDC 0.30 ~ 0.25 A 3% 10W 0.3W
HPU10D-111 40 ~ 50 VDC 0.25 ~ 0.20 A 3% 10W 0.3W
HPU10D-112 51 ~ 55 VDC 0.19 ~ 018 A 3% 10W 0.3W

HPU10D-101~105 are required to use AWG#20 / 4FT output cable.
HPU10D-106~112 are required to use AWG#24 / 4FT output cable.
The will be changed by modified output cable. electrical characteristics