HPU10B SERIES |
|
HPU10B SERIES
10W External power
supplies for medical applications including
home healthcare equipment,
patient monitors, blood pressure equipment,
portable medical devices, ECG machines, many
others
2
Year warranty.
FEATURES |
|
Wide
Operating Voltage, 80 to
275 VAC, 47 to 440 Hz |
|
UK Plug |
|
Medical
Safety 3rd (IEC60601-1
3rd Edition) |
|
Input to
Output : 2MOPP |
|
Single
Output |
|
Energy
Star 2.0, Efficiency
Level V |
|
Over
Voltage and Over Load
Protection |
|
Class II
System |
|
|
Electrical
Characteristics:
Vins
|
Safety Approvals Input Voltage Range
|
Safety Approval & Specification in Label |
100-240VAC |
Vin |
Operate Voltage Range
|
See derating curve |
80-275VAC |
Fi |
Input Frequency
|
Sine wave |
47-440Hz |
Po |
Output Power Range
|
See rating chart |
10W |
lil |
Input Current (Low Line)
|
lo=Full load, Vin=100VAC |
0.3A |
lih |
Input Current (High Line)
|
lo=Full load, Vin=240VAC |
0.2A |
Ir |
Low Line inrush Current
|
lo=Full load, 25°C, Cool start, Vin=100VAC
|
40A (max.) |
High Line Inrush Current
|
lo=Full load, 25°C, Cool start, Vin=240VAC
|
100A (max.) |
Eff |
Efficiency
|
lo=Full Load, Vin=230VAC |
67-85% |
REG-i |
Line Regulation
|
lo=Full Load, Vin=100-240VAC |
1% (mAX.) |
REG-o |
Load Regulation
|
Vin=230VAC |
3-5% |
OVP |
Over Voltage Protection
| |
Nil |
OCP |
Over Current Protection
|
Nil.But,Output protected to short circuit conditions
| |
Ttr |
Time of Transient Response
|
lo=Full Load, Vin=110VAC |
4mS (max.) |
Thu |
Hold-up Time
|
lo=Full Load, Vin=110VAC |
12mS (typ.) |
Ts |
Start Up Time
|
lo=Full Load, Vin=100-240VAC |
2S (max.) |
VP-p |
Ripple & Noise (Peak to Peak)
|
Full Load, Vin=90VAC |
1% (max.) |
Tc |
Temperature Coefficient
|
All Condition |
±0.04%/°C |
Pno |
No-Load Power Consumption
|
No load, Vin=230VAC |
See rating chart |
Vps |
Dialechic Withstanding Voltage
|
Primary to secondary |
6612VDC (min.) |
EMI |
EMC Emission
|
Compliance to EN55011 (CISPR11), EN61000-3-2,-3
|
B Class |
Environmental:
To | Operating Temperature | See derating curve | -10-70°C |
Ts | Storage Temperature | 10 — 95% RH | -40-85°C |
Ho | Operating Humidity | Non-condensing | 0-95% RH |
Hr | Storage Humidity |
See Rating Chart | 0-95% RH |
Vsg | Surge Voltage | All Condition | 2kV (min.) |
ESDa | Electro Static Discharge | Air Discharge, IEC61000-4-2 | 8kV (min.) |
ESDc | Electro Static Discharge | Contact Discharge, IEC61000-4-2 | 6kV (min.) |
MTBF | Mean Time Between Failure | Operating temperature at 25°C, Calculated per MIL-HDBK-217F | 0.1M Hrs (min.) |
ELEV | Operating Altitude (Elevation) | All condition | 5000m (max.) |
VBR | Vibration | 10 ~ 500Hz, 10min./1cycle,60min. Each along x, Y, Z axes. | 5G (min.) |
|
Mechanical
Specifications:
|
|
|
|
|